SKYLINE INSTRUMENTS CO.,LTD

Skyline Instruments Co.,LTD We only manufacture Lab Testing Equipment

Manufacturer from China
Verified Supplier
8 Years
Home / Products / Building Materials Flammability Tester /

Building Material Flammability Tester Ignitability Testing Equipment

Contact Now
SKYLINE INSTRUMENTS CO.,LTD
Visit Website
City:dongguan
Province/State:guangdong
Country/Region:china
Contact Person:MsIvy Xie
Contact Now

Building Material Flammability Tester Ignitability Testing Equipment

Ask Latest Price
Brand Name :SKYLINE
Model Number :SL-FL089
Certification :CE
Place of Origin :China
MOQ :1 unit
Price :negotiated
Payment Terms :T/T, L/C, Western Union
Supply Ability :1 unit/ month
Delivery Time :10 work days
Packaging Details :plywood case,
Building Material :Flammability Testing Equipment Ignitability Tester
Standards :EN ISO 11925-2, DIN 4102
Application :building products
Burner angle :45°
Materail :Stainless Steel
Anemometer accuracy :±0.1m/s.
Observing window :two big viewing windows on the front and right side
Timer accuracy :≤1s/h
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Building Material Flammability Testing Equipment Ignitability Test Single Flame Source

Description

Ignitability test, also called single-flame source test.Test apparatus corresponds to standard EN ISO 11925-2. Test results of the building structure can be classified according to standard 13501-1.

The dimensions of the test specimens shall be (250 ) mm long by (90 ) mm wide.

45°-tilted burner is advanced horizontally until the flame (length 20 mm) reaches the pre-set contact point with the test specimen. Tests may need to be conducted to either surface exposure or edge exposure, or both. The flame application time is 15 or 30 seconds as required by sponsor.

During the test following is recorded: whether ignition occurs; whether the flame tip reaches 150 mm above the flame application point, and the time at which this occurs; whether ignition of the filter paper occurs; observations of physical behavior of the test specimen.

Standards

EN ISO 11925-2, DIN 4102

Specifications

Power supply AC220V±10%,50Hz
Burner diameter 0.17mm burner with regulator; four ¢4mm air suction inlet
Gas source >95% propane gas;(user should provide for themselves)
Gas pressure 10kpa~50kpa
Flue air velocity 0.7m/s only on condition of burner burning and draught hood open
Time adjustment of continuous flame 0 ~ 99minutes99seconds
Time adjustment of afterflame 0 ~ 99minutes99seconds
Timer accuracy ±0.1m/s
Burner angle 45°
Material stainless steel
Observing window two big viewing windows on the front and right side

Anemometer accuracy

±0.1m/s

Whole set: control device, burning chamber,burner,magnetic valve,flame lighter, signal control line and specimen holder.

Equipped with a digital anemometer/thermometer and a stopwatch for simple but accurate measurement of the flow, temperature and time.

Building Material Flammability Tester  Ignitability Testing Equipment

Inquiry Cart 0