SKYLINE INSTRUMENTS CO.,LTD

Skyline Instruments Co.,LTD We only manufacture Lab Testing Equipment

Manufacturer from China
Verified Supplier
9 Years
Home / Products / Fire Testing Equipment /

Fire Resistance Tester for Gypsum Board Flammability stability tester for Paper Gypsum Board

Contact Now
SKYLINE INSTRUMENTS CO.,LTD
Visit Website
City:dongguan
Province/State:guangdong
Country/Region:china
Contact Person:MsIvy Xie
Contact Now

Fire Resistance Tester for Gypsum Board Flammability stability tester for Paper Gypsum Board

Ask Latest Price
Video Channel
Brand Name :SKYLINE
Model Number :SL-SGF01
Certification :CE
Place of Origin :China
MOQ :1
Price :negoitable
Payment Terms :T/T
Supply Ability :1 unit/Month
Delivery Time :15 working days
Packaging Details :Plywood case
Power :220V±10%50Hz/180W
Lighting :Autoignition
Timing accuracy (resolution) :0.1 seconds
Temperature stabilization time :<3 minutes (800±30) ℃
The flow regulation range :100 to 1000 ml /Min
Flame length :20mm to 100mm(Adjustable)
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Application

It is used for mensurating plasterboard stability in case of fire, can also be applied to other building materials, decoration, decoration materials, fire prevention performance measurement.

Test Introduction

It is introduced drilling and processed under prescribed conditions. Suspend the sample between the two fire nozzles with a support rod, making the fire nozzles perpendicular to the surface of the sample. Liquefied petroleum gas is used as the heat source to supply gas to the two burners of the fire stability tester. The distance between the burner nozzle and the plate surface is 30mm. Suspend the load at the lower end of the specimen as per the standard and ignite the burner. Measure the temperature at a distance of 5mm from the plate surface using a thermocouple. At the initial stage of the test, the carbonized protective paper that has fallen onto the thermocouple should be removed without shaking the sample. By adjustment, the temperature was controlled at (800±30) ℃ within 3 minutes and maintained at this temperature throughout the test. Start timing from the moment the specimen is exposed to fire until it breaks and fails. Record the time it takes for each specimen to be burned off, and take the minimum value among the five specimens as the fire stability of this group of specimens, accurate to 1 minute.

Test Parameter

1.Supply voltage: 220W + 10%, 50HZ

2.Power: 180W

3.Ignition:Automatic ignition

4.Thermocouple:Double thermocouple temperature measurement

5.Timer:temperature and time digital display

6.Burner:Double burner, for Φ 9.5 mm + / - 0.5 mm, inner diameter is about 100 mm long, there is air conditioning nut 10 mm, spray fire head (40 + 1) mm in diameter, Spray fire flame hole diameter (2.5 + 0.1) mm

7.Fire Height:20 mm - 100mm

8.Burning duration:adjustable applying time and duration of a burning flame in 0 ~ 99 minutes, 99 seconds range adjustment test process automatic control.

9.The style break fall automatically record the combustion time.

10.Source: test flames of air using liquefied gas or oil gas (suggest try choosing temperament better gas)

11.Flow adjusting range is 100 ~ 1000 ml/Min

12.Timing precision (resolution) : 0.1 s

13.Test temperature: 800-30 ℃, the highest temperature 900 ℃, precision grade 1 time temperature stability: < 3 min (800 +) 30 ℃

14.Specimen: 300 mm (length) x 50 mm (width), on both sides of the distance of 25 mm hole

15.Hanging load respectively: 7, 10, 12, 15, 17, 20 n (0.75 ~ 2.5 ㎏ conversion unit suspended load)

16.Dimension 700 (width) x 300 (deep) x 600 (high) : (mm) 800 mm outer

17.Packing size (width) x 420 (deep) x 700 (high)

Test Standard

GB/T9775-2008(ISO6308:1980)

Installation Sheet

Place Dimension 700(W)×300(D)×600(H)mm
Weight About 30
Test table Require no less than 1 meter and a width of no less than 0.75 meters
Power Supply vlotage 220V±10% ,50HZ
Power 180W
Water Source NA
Air Source Liquefied petroleum gas
(It is recommended to choose gas with better quality as much as possible)
Pollutants Dust
Ventilation requirements The equipment needs to be placed in a fume hood or collectedfor flue gas and connected to the flue gas treatment and purification system
Other test requirement Gas cylinder matching pressure reducing valve (capable of adjusting to 0.3 megapascals)
Inquiry Cart 0